Continuous Glucose Monitoring in the Management of Neonatal Hypoglycaemia
McGlacken-Byrne, S.M; O’Neill, R.; Jenkinson, A.; Murphy, J.F.A.
A quality improvement project was carried out in a Level 3 Neonatal Intensive Care Unit (NICU) which aimed to successfully implement the use of continuous glucose monitoring systems (CGMS) in hypoglycaemic infants. Piloting of the device revealed several potential practical barriers to its reliably successful implementation. Five Plan-Do-Study-Act (PDSA) cycles followed, tackling these problems and other issues inductively identified throughout the project. Parents and multi-professional stakeholders were involved and consulted throughout. Change was measured on a runchart using qualitative and quantitative feedback. Problem rate per patient was reduced to zero by the end of a one-month study period. This study used basic quality improvement methodologies to implement a change intervention in a structured manner and elucidated aspects of its use that need to be adapted for its successful incorporation into real-life clinical practice.
↧